CO2-O2-MODULE
The CO2-O2-MODULE (included automatically with purchase of LEO 2.0) measures the CO2 and O2 % concentration in the incubator’s sampled gas. The measured values are temperature- and pressure-compensated. CO2-O2-MODULE’s lifetime is 5 years.
Okolab recommends once-yearly calibration. No need to part ways with your meter to keep the sensor calibrated: CO2-O2-MODULE can be easily extracted from LEO 2.0 (and immediately replaced with a spare, if available) and sent out for factory re-calibration. We offer a worry-free yearly sensor replacement service, OKO-CARE, where you automatically receive a newly-calibrated sensor once per year.
Features – CO2
- Sensor: Non Dispersive InfraRed detector (NDIR)
- Measurement range: 0-20%
- Accuracy (in the range 0-10%): ±(1% of Full Scale +2% reading)
Features – O2
- Sensor: Fluorescence-based optical sensor
- Measurement range: 0-22%
- Accuracy (in the range 0-10%*): ±(1% of Full Scale +2% reading)
*Measurement range 0-10% is guaranteed for 5 years