H303-T-H-CO2/O2 [0-21]

Bold Line 3 – Temperature, Humidity and CO2/O2 controller.

The bundle includes:

  • temperature controller for BL3 electric chambers
  • active humidity
  • CO2 and O2 controller
  • air pump
  • user touch screen interface

 

The temperature module:

  • Allows to work in chamber or in specimen feedback mode and to continuously monitor room temperature.
  • T range: 3°C above ambient – 60°C.
  • Specimen Temperature accuracy: ± 0.1°C in specimen feedback mode, ± 0.3°C in chamber feedback mode, if room temperature remains within ± 1°C.
  • Automatic self-calibration routines for the incubating chamber and for the objective heater.

 

The CO2 and O2 module:

  • CO₂ and O₂ concentrations are digitally controlled by mixing Air (from the Okolab air pump), N₂ and CO₂.
  • CO₂ range: 0–20%
  • O₂ range: 0–21%
  • CO₂ sensor: drift-free Non Dispersive InfraRed (NDIR) dual wavelength detector (10 years lifetime, accuracy ± 0.1%).
  • O₂ sensor: optical oxygen sensor (expected lifetime 5 years, resolution 0.1%, accuracy ± 0.1% at 5% O₂).
  • Total output flow rate is variable in the range: 0.2 – 0.4 l/min. (Factory setting: 0.4 l/min.)

 

The Active Humidity Controller:

  • Humidity sensor to control gas relative humidity.
  • Operating thermal range: 25÷50°C.
  • Humidity range depends on working temperature. At 37°C humidity can be regulated in the range 50 to 95%.
  • Humidity Set-point Resolution: 1%.
  • It features a temperature controlled tube to deliver the humidified gas to the incubating chamber.

 

Integrated into LASX 2.0, NIS-Elements, MetaMorph and MicroManager Imaging Software and capable to receive TTL input. Macro for ZEN Black software available.

Compatible with the CO2-H-IN-CHAMBER SENSOR for in chamber closed loop control on CO2 and Humidity.